Key Takeaway

Platinum Health Recalls LumaRail Adult Portable Bed Rails Due to Serious Entrapment and Asphyxia Hazard; One Death Reported was recalled on February 23, 2023. What to do: Consumers should immediately stop using the recalled bed rails and contact Platinum Health for a free repair kit. Consumers who use model PHB7000D on...

CPSC Ongoing

Platinum Health Recalls LumaRail Adult Portable Bed Rails Due to Serious Entrapment and Asphyxia Hazard; One Death Reported

Recalled: February 23, 2023 ~About 53,000 units affected 23136

Description

This recall involves the Bed Assist Rail with Legs (model PHB4000), Free Standing Bed Assist Rail (model PHB6000) and the DoubleSafe Bed Assist Rail (model PHB7000D). The bed rails are made of white metal tubing with a gray or black foam rubber grip handle and may have a black fabric pouch. The name "Platinum Health" is printed on a label located on the leg of the bed rail or on the grip handle. The bed rails also have a light mount below the grip handle.

Injuries / Consequence

Platinum has received one report of a death involving the Bed Assist Rail with Legs (model PHB4000). In October 2021, an 81-year-old man at a nursing home in Pennsylvania died from positional asphyxia after becoming entrapped between his mattress and the bed rail.

Remedy

Consumers should immediately stop using the recalled bed rails and contact Platinum Health for a free repair kit. Consumers who use model PHB7000D on twin beds will receive a free replacement bedrail redesigned to properly fit twin beds.

Products Affected

LumaRail Bed Assist Rails

Images

Platinum Health Recalls LumaRail Adult Portable Bed Rails Due to Serious Entrapment and Asphyxia Hazard; One Death Reported Platinum Health Recalls LumaRail Adult Portable Bed Rails Due to Serious Entrapment and Asphyxia Hazard; One Death Reported Platinum Health Recalls LumaRail Adult Portable Bed Rails Due to Serious Entrapment and Asphyxia Hazard; One Death Reported Platinum Health Recalls LumaRail Adult Portable Bed Rails Due to Serious Entrapment and Asphyxia Hazard; One Death Reported

Frequently Asked Questions

Is this recall still active?

Check the status badge at the top of this page. "Ongoing" means the recall is still active and you should follow the remedy instructions. "Completed" or "Terminated" means the recall process has concluded, though the safety issue remains relevant for any unrepaired products.

How do I check if my product is affected?

Compare the model number, lot code, serial number, or date range listed in the recall notice with the information on your product's label or packaging. For vehicles, check your VIN at NHTSA.gov/Recalls.

What remedy is available for this recall?

The listed remedy for this recall is: Consumers should immediately stop using the recalled bed rails and contact Platinum Health for a free repair kit. Consumers who use model PHB7000D on twin beds will receive a free replacement bedrail...

Where can I find the original recall notice?

This recall was issued by CPSC. You can view the original notice using the link provided above.